Dark Light
-49%
,

Conductivity Transmitter

Availability:

6 in stock


  • Direct change over to Conductivity (µScm )
  • TDS measurement (ppm )
  • Automatic temperature compensation
  • 4-20 mA Isolated Output

22,990.00  45,500.00 

6 in stock

*
Compare

Product Description

Conductivity/Hardness/Resissvity Online analyzer, an intelligent Online chemical analyzer, is widely applied for corneous monitoring and measurement of EC value or TDS value or ER value and temperature in the solution in the industry of thermal power, chemical fernier, environmental parecon, metallurgy, pharmacy, biochemistry, food and water, etc.
The model WKEC100 is used for the conductive measurement/control of electrrolyc conductive, resissisvity or the value.
Conductivity is a function of ion on centration, ionic c get, and ion mobility.
Ions in water conduct current when an electrical potential is applied across electrodes immersed in the
solution. A controller system consists of a microprocessor-based controller and a conductivity probe. 3 Electrode cells(K=0.01,0.1 and 1.0)can be connected to the device. Temperature serves as the second input variable, measured by a NTC10K/ PT1000 probe. Depending on the measured variable, it is therefore possible to implement specific, automatic temperature compensation. All adjustments to the current outputs, alarm relays, and calibration of the conductivity and temperature inputs can be made using the
controller’s .

Technical data

Technical Specification

Power supply

AC:220VAC±10% or 110VAC 50Hz/60Hz
DC:24VDC±20% Input power≥6W

Measuring range

0.00~2000µS/cm(max.20000µS/cm)

Temperature range

-10~130℃

Serial communications

Rs485

Output

4-20 mA + relay

EC/TDS/Resisvity

±1%FS

NTC10K

±0.3℃

Pt1000

±0.3℃

Relay humidity

5 ~ 95%RH(No condensaon)

Operating Temperature

0℃~60℃

Storage

-15℃~ 65℃

Screen size

2.8inch

Dimension

Overall dimension: 100mm*100mm*150mm(H*W*D)
Cutout dimension:92.5mm*92.5mm(H*W)

Weight

0.65Kg

Ingress protecon

IP 54

Type

NTC10K/PT1000

Model

Manual/automatic

Output

Overall dimension: 100mm*100mm*150mm(H*W*D)
Cutout dimension:92.5mm*92.5mm(H*W)

Relay

2 relays Ac250

Downloads

Datasheet : Conductivity Transmitter

 

Delivery Scope

  • To achieve the best possible accuracy, the calibration should cover the range of the desired measurement values. If the readings go beyond the calibrated range, the pH meter assumes linearity and simply extrapolates the value to be displayed. The true value may be slightly different. More advanced pH meters will let the user calibrate at three, four or five and even higher numbers of pH values. A multi-point calibration mean, in comparison to a two-point calibration, that you can calibrate your pH tester on both sides of the zero point (pH 7.00). This will expand your pH .measurement range without the need of re-calibrating.